Special Features:-
We are leading exporter of Microprocessor Nephelo Turbidity Meter. Errors & Precautions: Start : 1- Connect Adopter connection on rear panel & switch “ON” the Instrument. Adopt the Calibration procedure, only if needed and when You have proper standards . Otherwise the Instrument is Pre-calibrated with default values ( Factory Settings). Turbidity sample test tube must be thoroughly cleaned both Inside and outside. In case the test tube gets scratched or etched discard it. Do not touch the Turbidity test tube where the light strikes.a. at side of test tube. So hold test tube at the top. Please ensure that the mark on the Turbidity test tube Coincides with the mark on the Turbidity sampler & fill the Sampler tube up to the mark given.
Technical Specifications:-
Model | S- 968 |
Range | 0-400 NTU (in 5 Ranges), 0-2, 10, 100, 200, 400 |
Resolution | 1 NTU |
Accuracy | ±/- 3% FS, ±1 Digit |
Display | 4 Digit 7 Segment LED |
Light Source | 3 V, 0.3 Amp LED Lamp |
Detector | Photodiode |
Sample System | 30 mm Clear Glass Test Tube |
Power Supply | 230V, +-10% AC, 50 Hz |
Dimension | 195X230X95 mm (LXBXH) |
Weight | 3 KG |